![China Electronics Technology Instruments AV3672 Series Скачать руководство пользователя страница 330](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790330.webp)
11 Technical Indicators and Measurement Methods
11.2 Measurement Methods
326
Test of effective direction of Port 2
21) Connect the load to the test port 2. Click [Analysis] > [Memory] > [Trace Operation] > [Close].
22) Change the start and stop value of [User Range], and search the maximum value within the set range. The
maximum value is the effective direction value of the test port 2.
23) Refer to the full double-port calibration steps of the test port 1 and 2, i.e. Step (5) to (9), and calibrate Port 3
and 4.
Test of transmission tracking of Port 3
24) Click [Meas] > [S34]. As shown in Fig. 13, reconnect the test cable to Port 3 and 4.
25) Change the start and stop value of [User Range]. Search the “Max” and “Min” within the set range. The larger
absolute value is the transmission tracking value of the test port 3.
Test of transmission tracking of Port 4
26) Click [Meas] > [S43].
27) Change the start and stop value of [User Range]. Search the “Max” and “Min” within the set range. The larger
absolute value is the transmission tracking value of the test port 4.
Effective load matching of test port 3
28) Click [Meas] > [S33].
29) Change the start and stop value of [User Range]. Search the maximum value within the range. The maximum
value is the effective load matching value of Port 3.
Test of effective load matching of Port 4
30) Click [Meas] > [S44].
31) Change the start and stop value of [User Range]. Search the maximum value within the range. The maximum
value is the effective load matching value of Port 4.
Test of reflection tracking of Port 3
32) Connect the short-circuit device to the test cable end face of the test port 1. Click [Meas] > [S33]. Then click
[Analysis] > [Memory] > [Normalize].
33) Change the start and stop value of [User Range]. Search the “Max” and “Min” within the set range. The larger
absolute value is the reflection tracking value of the test port 3.
Test of effective source matching of Port 3
34) Click [Analysis] > [Memory] > [Trace Operation] > [Data + memory].
35) Connect the open-circuit device to the test cable end face of the test port 3. Click [Scale] > [Reference value],
and enter -50.
36) Change the start and stop value of [User Range]. Search the maximum value within the set range. The effective
source matching value of the test port 3 is the difference of the obtained marker value minus 6.
Test of effective direction of Port 3
37) Connect the load to the test port 3. Click [Analysis] > [Memory] > [Trace Operation] > [Close].
38) Change the start and stop value of [User Range], and search the maximum value within the set range. The
maximum value is the effective direction value of the test port 3.
Test of reflection tracking of Port 4
39) Connect the short-circuit device to the test cable end face of the test port 4. Click [Meas] > [S44]. Then click
[Analysis] > [Memory] > [Normalize].
40) Change the start and stop value of [User Range]. Search the “Max” and “Min” within the set range. The larger
absolute value is the reflection tracking value of the test port 4.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...