11 Technical Indicators and Measurement Methods
11.2 Measurement Methods
317
Test the frequency resolution indicators of frequency sweep signals output by AV3672 series vector network
analyzers.
11.2.2.2 Instrument and Accessories for Test of Frequency Resolution
1)
spectrum analyzer
AV4038
2)
BNC cable
1.
3)
RF cable
1.
11.2.2.3 Steps of Frequency Resolution Test
1) As shown in Fig. 11.2, connect the system and preheat is for 60min at least.
2) Press [Reset] to set the initial state of AV3672 series vector network analyzers.
3) Press [Sweep] > [Sweep type] > [CW] > [Frequency], and enter 10GHz. Press [Stimulus] > [Trigger] > [Hold].
Record the measurement result of the spectrum analyzer.
4) Change the center frequency by 1Hz. If the reading of the spectrum analyzer also changes by 1Hz, the frequency
resolution conforms to the technical indicator requirements.
Fig. 11.2 Block Diagram of Frequency Resolution Test
11.2.3 Test of Port Output Harmonic Suppression
11.2.3.1 Items of Test of Port Output Harmonic Suppression
Test the harmonic suppression indicators of frequency sweep signals output by the port of AV3672 series vector
network analyzers.
11.2.3.2 Instrument and Accessories for Test of Port Output Harmonic Suppression
1)
spectrum analyzer
AV4038 or R&SFSUP50
2)
RF cable
1.
11.2.3.3 Steps of Test of Port Output Harmonic Suppression
1) As shown in Fig. 11.3, connect the system and preheat is for 60min at least.
2) Press [Reset] to set the initial state of AV3672 series vector network analyzers. Click [Channel] > [Hardware
configuration] > [Path configuration]. Select the filtering mode for the low-frequency segment of Port 1 and 3.
Press [Power] > [-5] and the ENTER key.
3) Press [Sweep] > [Sweep Time] > [10] and the ENTER key.
4) Press [Sweep] > [Sweep type] > [CW] > [Frequency], and enter 10MHz. Set the corresponding test
frequency of the spectrum analyzer and record the reading of the spectrum analyzer. Set the second
harmonic frequency of the spectrum analyzer as 20MHz and record the reading of the spectrum analyzer.
The difference of two readings is the port harmonic suppression indicator. Get difference between reading
of second harmonic and reading of main signal. Indicators of AV3672A/B are divided into two segments. The
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...