![China Electronics Technology Instruments AV3672 Series Скачать руководство пользователя страница 260](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790260.webp)
8 Basis of Network Measurement
8.5 Group Delay
256
Relationship of the aperture and group delay:
If the aperture is too wide
and the phase shift of adjacent frequency points is more than 180°, errors will occur in group
delay calculation.
2) Apply the smoothing function of the analyzer.
Perform sweep once, enable the smoothing function and change the value percentage in smoothing settings. This is
similar to change of the frequency interval between sweep points. This method allows a wider group delay aperture,
and the phase shift is allowed to exceed 180° within the smoothing aperture.
Group delay measurement can be done in the following sweep types:
a) Linear frequency;
b) Logarithmic frequency;
c) Segment sweep.
The group delay aperture depends on the frequency interval and the sweep point density, therefore, it changes in the
logarithmic frequency and segment sweep mode.
8.5.4 Improvement of Measurement Accuracy of Group Delay
It is important to keep the phase difference of adjacent measurement points less than180° during group delay
measurement; otherwise, the phase and group delay information will be incorrect, as shown in Fig. 8.15.
Insufficient sampling may occur in measurement of the device of long electrical delay. Adjust the following
settings until the measurement trace does not change so as to ensure that the phase difference of two adjacent points
is less than 180°.
1) Increase the number of points.
2) Reduce the frequency span.
Fig. 8.15 Influence of Insufficient Sampling on Group Delay Measurement
The frequency response is a main error of group delay measurement. This error can be significantly reduced by
means of through type response calibration. Perform the full double-port calibration to realize higher accuracy. For
the amplifier, the response may vary with the temperature. Therefore, the test should be done at the working
temperature of the amplifier.
8.5.5 Steps of Group Delay Measurement
1) Reset the analyzer. Adjust the source power of the analyzer to test the amplifier.
a) Set the source power of the analyzer within the linear zone of the analyzer response (generally at least 10dB less
than the 1dB input compression point).
b) If required, connect one attenuator to the output port of the amplifier to fully attenuate the output power of the
amplifier and prevent the receiver at the reception end of the analyzer from compression or burning.
2) Connect the DUT, as shown in Fig. 8.16.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...