9 Remote control
9.2 Programmed Port and Configuration of Instrument
279
1) Initialize the instrument state before changing the setting.
For instrument setting of remote control, initialize the instrument state (sending “*RST”, etc.) and then set the
required state.
2) Command sequence
Generally, the setting command and query command should be transmitted separately. Otherwise, the returned
value of the query command will change according to the current instrument operation sequence.
3) Fault response
The service request must be sent by the instrument. The controller of the test system should be programmed to
guide the instrument to actively send the service request if required, so as to enable the corresponding service
interruption program.
4) Error queue
The error queue instead of the state register should be inquired once the service request is processed by the
controller program, so as to find more accurate causes. The queue should be always inquired in the controller
program test phase to obtain the commands sent by the controller to the instrument.
9.2 Programmed Port and Configuration of Instrument
●
LAN
●
GPIB
9.2.1 LAN
The programmed LAN (Local Area Network) system is configured with SICL-LAN to control AV1443 series
vector network analyzer.
Attention
Use of USB master control port connector of the front panel:
The Type-A connector of the front panel is a USB master control port connector. In the AV1443 series vector
network analyzer, this port is connected to the flash disc with USB1.1 interface to upgrade the software in the
instrument. It can also be connected to the USB keyboard and mouse to control the vector network analyzer. It must
not be used for program control of the instrument.
9.2.1.1 Establish Connection
Connect the AV3672 series vector network analyzer and external controller (computer) to LAN through network
cables, as shown in Fig. 9.4.
Fig. 9.4 Connection Diagram of LAN Interface
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...