8 Basis of Network Measurement
8.2 Phase Measurement
247
Similar to the amplitude measurement, the S-parameter is also applied in phase measurement. Phase measurement
is a kind of relative (ratio) measurement instead of absolute measurement. The phase of the signal into the device
(incident signal) and the response signal are compared in phase measurement. The response signal may be reflected
signal or transmitted signal. Given that the analyzer is accurately calibrated, the phase difference (i.e. phase shift)
of two signals is the measurement result of phase characteristics of the DUT. The following figure shows the phase
deviation between the incident signal and transmitted signal observed through the oscilloscope.
Fig. 8.7 Phase Shift between Signals
8.2.2 Why to perform phase measurement?
Phase measurement is an important function of the vector network analyzer. Fig. 8.8 shows the reasons for accurate
measurement of the amplitude and phase.
Fig. 8.8 Reasons for Accurate Amplitude and Phase Measurement
If the element and circuit are used to transmit signals in the communication system, signal distortion must be within
the normal limits. The signal distortion can be divided into two types.
1) Linear distortion: refer to the failure to keep the flat amplitude and linear phase shift within the relevant
frequency range as a result of frequency changes.
2) Nonlinear distortion: refer to the new spectrum component of the circuit, such as AM-PM transformation.
It is important to accurately measure the amplitude and phase characteristics of the element or circuit, so as to
ensure the efficient transmission or adsorption of energy in the circuit and prevent the distortion in signal
transmission. A good example to prevent distortion is to measure the complex impedance of the antenna.
8.2.3 Use of Phase Format of Analyzer
The phase format of the analyzer is used to display the phase changes over the frequency or power. The phase
difference beyond ±180° between the reference signal and response signal cannot be measured by the analyzer. If
the phase value varies from +180° to -180°, the sawtooth type phase measurement trace will be displayed on the
analyzer. The sawtooth waveform may not reach +180° or -180° sometimes in measurement at the discrete
frequency points. The data points of +180° and -180° may not be the points of sweep measurement.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...