Page 9
4.2 Voltage Inputs
When measuring voltage, each of the two inputs can be set to each of the four
gain ranges as below:
Configuration
Number
Type Code
Type
Pri Range Sub Range Excitation
1
03
Volts
78mV
78mV
(N/A)
2
N/A
Volts
78mV
300mV
(N/A)
3
04
Volts
78mV
5V
(N/A)
4
N/A
Volts
78mV
10V
(N/A)
5
N/A
Volts
300mV
78mV
(N/A)
6
N/A
Volts
300mV
300mV
(N/A)
7
N/A
Volts
300mV
5V
(N/A)
8
N/A
Volts
300mV
10V
(N/A)
9
05 or 00
Volts
5V
78mV
(N/A)
10
01 or 00
Volts
5V
300mV
(N/A)
11
06 or 00
Volts
5V
5V
(N/A)
12
00
Volts
5V
10V
(N/A)
13
N/A
Volts
10V
78mV
(N/A)
14
N/A
Volts
10V
300mV
(N/A)
15
N/A
Volts
10V
5V
(N/A)
16
N/A
Volts
10V
10V
(N/A)
NOTE : The ‘Type Code’ is a value which can be used as a shortcut to configure
channels and matches the codes used within the NetScanner® 9146-R
configuration.
Содержание FlightDaq-TL
Страница 20: ...Page 19 5 2 The Setup tab Figure 5 21 Main Setup page...
Страница 25: ...Page 24 Figure 5 52 Live Data Page Figure 5 53 Live Data Page...
Страница 46: ...Page 45 Select an additional Scan List in this case we have chosen 4...
Страница 47: ...Page 46 Then disable the secondary and tertiary data streams...