Page 48
7 Service and Calibration
7.1 Service
There are no user serviceable parts inside the instruments. Should any
difficulties be encountered in the use of the FlightDaq-TL, it is recommended
that you contact Chell Instruments Ltd for advice and instructions.
7.2 Calibration
Calibration is recommended on an annual basis and Chell Instruments Ltd.
provides a fully traceable facility for this purpose.
7.3 Adjustment
7.3.1
Changing the Power Mode of the device
1.
Remove all fourteen screws from the baseplate using a 2.0mm hex driver.
2.
Carefully remove the baseplate from the device to avoid damaging the gasket.
3.
Adjust DIP switch to required settings. See table below.
4.
Change the metal identifying tag on the side of the case to show the correct mode
(DC or PoE).
5.
Re-insert the gasket in the correct position and screw down baseplate using the
same hardware.
Internal DIP Switch Settings for flightDAQ-TL
SW1 SW2 SW3 Mode
OFF OFF OFF Normal operation - PoE Mode
(Default)
ON
ON
ON
Normal operation - DC Mode
ON
OFF -
Hard reset function
OFF ON
-
Factory default function
7.3.2
Factory default procedure
1.
Power off unit.
2.
Change DIP switch to Factory default function (SW3 is either ON or OFF).
3.
Power on unit.
4.
Wait for Red/Blue LEDs to alternate flash.
5.
Power off unit.
6.
Change DIP switch back to normal operation settings.
Default setting showing
all switches off
Содержание FlightDaq-TL
Страница 20: ...Page 19 5 2 The Setup tab Figure 5 21 Main Setup page...
Страница 25: ...Page 24 Figure 5 52 Live Data Page Figure 5 53 Live Data Page...
Страница 46: ...Page 45 Select an additional Scan List in this case we have chosen 4...
Страница 47: ...Page 46 Then disable the secondary and tertiary data streams...