background image

Page 27 

 
 

‘Transducer Variant’ 

Here the variant of transducer is selected (when 
applicable). The options in this menu are dependent on 
the ‘Transducer Type’ selected. 

‘Unit String’ 

Here the units for the measurement can be selected. This 
is a string value and as such, it is a free text field. 

‘Serial’ 

Here the serial number of the transducer can be entered 
(if applicable). 

‘Chell Part Number’ 

Here the Chell part number of the transducer can be 
entered (if applicable). 

‘User Part Number’ 

Here the User part number of the transducer can be 
entered (if applicable). 

‘Primary FSV’ 

Here the full scale of the primary input is configured. If 
you have selected a specific transducer type (such as 
RTD), then these fields will be fixed. If, however, you 
have selected Voltage as a ‘Transducer Type’ then the 
following range options will be available. 

 

‘Secondary FSV’ 

Here the full scale of the secondary input is configured. If 
you have selected a specific transducer type (such as 
RTD), then these fields will be fixed. If, however, you 
have selected Voltage as a ‘Transducer Type’ then the 
following range options will be available. 

 

‘User Span’ 

Here a calibration coefficient can be entered for this 
channel. The calibration takes the form of y=mx+c and 
the span value corresponds to 

m.

 

‘User Zero’ 

Here a calibration coefficient can be entered for this 
channel. The calibration takes the form of y=mx+c and 
the span value corresponds to 

c.

 

‘Apply Settings’ 

Once the configuration is complete, click on apply 
settings to save the configuration. The unit must be soft-
restarted to adopt these settings. If you wish all channels 
to be the same, make sure that ‘Configure for all 
channels’ is selected before clicking in ‘Apply Settings’ 

 
 
 

 

Содержание FlightDaq-TL

Страница 1: ...lgate Road North Walsham Norfolk NR28 0AJ ENGLAND Tel 01692 500555 Fax 01692 500088 FlightDaq TL Multi Functional Scanner INSTALLATION AND OPERATING MANUAL e mail info chell co uk Visit the Chell webs...

Страница 2: ...in this manual may impair the user s protection Chell Document No 900232 Issue 1 5 ECO 4559 Date 13th September 2021 Chell s policy of continuously updating and improving products means that this manu...

Страница 3: ...ction 18 5 2 The Setup tab 19 5 3 Common Controls Sidebar 20 5 4 The Setup Parameters Page 21 5 4 1 Introduction 21 5 4 2 Data Streaming 21 5 4 3 General Ethernet Comms 22 5 5 Live Data Page 23 5 7 Th...

Страница 4: ...Page 3 7 3 1 Changing the Power Mode of the device 48 7 3 2 Factory default procedure 48 7 4 Cleaning 49 Appendix 1 Channel Configurations 50...

Страница 5: ...tric pressure transducer The FlightDaq TL can be configured for any combination of channels without comprising speed or performance with the exception of the 24V excitation which must be applied to al...

Страница 6: ...A and B 2 2 Package Dimensions 241 2 x 89 x 68mm Weight 1 53 Kg 2 3 Ethernet Specifications TCP IP 10Mb s 100Mb s via Auto Negotiation TCP and UDP protocols supported 2 4 Operating conditions Operati...

Страница 7: ...e 1 Remove all fourteen screws from the baseplate using a 2 0mm hex driver 2 Carefully remove the baseplate from the device to avoid damaging the gasket 3 Adjust DIP switch to required settings See ta...

Страница 8: ...nection Type Lemo 1B series plug Part Number FXG 1B 308 XXXXXX plus strain relief Example part numer FDG 1B 308 CYBD52Z plus GMA 1B 045 RA strain relief Pin Number Function 1 Vs Secondary Input 2 24 V...

Страница 9: ...uts which can be configured for four different ranges 78mV 300mV 5V and 10V In addition excitation can be provided on pin 1 for resistance and ratiometric ranges and pin 2 for 24 volts excitation In a...

Страница 10: ...10V N A 5 N A Volts 300mV 78mV N A 6 N A Volts 300mV 300mV N A 7 N A Volts 300mV 5V N A 8 N A Volts 300mV 10V N A 9 05 or 00 Volts 5V 78mV N A 10 01 or 00 Volts 5V 300mV N A 11 06 or 00 Volts 5V 5V N...

Страница 11: ...et to 78mV to measure the hot junction and the secondary input is set to 5V which when combined with the excitation will measure the output of the thermistor The results from the two inputs are conver...

Страница 12: ...thermocouple is correct The heater is enabled via the web server and will apply a current to a heating resistor bonded to the cold junction for 15 seconds The user can then determine if the polarity...

Страница 13: ...figured to measure resistance over a number of ranges as in the table below Configuration Number Type Code Type Variant Range Sub Range Excitation 25 30 Resistance 250R 300mV 1 k 26 31 Resistance 500R...

Страница 14: ...s over a number of ranges as in the table below Conversion from resistance to temperature is carried out by using the in built NIST tables Configuration Number Type Code Type Variant Range Sec Range E...

Страница 15: ...aq TL can be configured to measure RTD 385 pt100 using the four wire technique to remove the effect of lead resistance as below Configuration Number Type Code Type Variant Range Pri Range Sec Range Ex...

Страница 16: ...fied The FlightDaq TL can be configured to measure a range of amplified pressure transducers These can have an excitation of 10V or 24V The FlightDaq TL can be configured for transducer output full sc...

Страница 17: ...Exc SPARE 78mV 24V 38 105 Pressure Amplified Sub 300mV 24V Exc SPARE 300mV 24V 39 106 Pressure Amplified Sub 5V 24V Exc SPARE 5V 24V 40 107 Pressure Amplified Sub 10V 24V Exc SPARE 10V 24V 41 108 Pres...

Страница 18: ...TL can be configured for transducer output full scales of 78mV 300mV 5V or 10V as below Configuration Number Type Code Type Variant Range Pri Range Sec Range Excitation 45 112 Pressure Ratiometric 78m...

Страница 19: ...as of functionality namely Setup Live Data Channel Configuration Advanced and Factory Tools Setup provides the means to set FlightDaq TL s main operating parameters such as data streaming rates and de...

Страница 20: ...Page 19 5 2 The Setup tab Figure 5 21 Main Setup page...

Страница 21: ...ions Control Function RESET button Resets the FlightDaq TL similar to power cycling the device Use to activate new settings and or rebuild calibration tables The option exists for a hard reset and a s...

Страница 22: ...e 5 41 Data Steaming Setup Control Function Comms Protocol radio button Chooses the communication protocol that is to be used This button changes what options are available below it Data Rate option l...

Страница 23: ...2 TCP Comms group IP Address IP address allocated to FlightDaq TL on the user s network Subnet Subnet mask as set on the user s network Local port Local port of the device fixed to 101 Apply TCP Appli...

Страница 24: ...layed on the page and updated at a rate of around 2Hz The type of value shown in the live data window may be selected by means of the option buttons in the righthand frame These are as follows Enginee...

Страница 25: ...Page 24 Figure 5 52 Live Data Page Figure 5 53 Live Data Page...

Страница 26: ...y tables The summary table appears at the top of the Channel Configuration tab and gives the user a quick summary of the current configuration of the primary channels as shown below in figure 5 72 Fig...

Страница 27: ...onfiguration data resides This will always be internal unless Smart Lemo s are used When a Smart Lemo is used the configuration including any look up table s are stored within the Smart Lemo Config he...

Страница 28: ...ge as a Transducer Type then the following range options will be available Secondary FSV Here the full scale of the secondary input is configured If you have selected a specific transducer type such a...

Страница 29: ...l will appear as shown in Figure 5 74 below Figure 5 74 Channel Configuration with Smart Lemo As can be seen all settings stored to and therefore read from the Smart Lemo are identified with an asteri...

Страница 30: ...ate Smart Lemo connected which enables the calibration for that transducer to be tied to and travel with that transducer 5 8 1 Standard In Built Look Up Tables The standard in built look up tables ava...

Страница 31: ...tab as below Figure 5 81 Channel Configuration group It can be seen that in this case channel 2 is set to a transducer type of Voltage Under these circumstances the internal LUT editor is available a...

Страница 32: ...values entered above into the LUT Remove last entry pair This will remove the last entry in the LUT 5 8 2 1 Loading a custom LUT from file A custom LUT can be loaded into the FlightDaq TL memory from...

Страница 33: ...elect an appropriate file Then select Import and the file will be loaded into the FlightDaq TL as below Figure 5 83 Look up table editor This shows that file is loaded into the memory for the primary...

Страница 34: ...red for a 5V input The LUT will convert voltage an engineering unit First enter the measured input value in this case volts and the output value in engineering units Figure 5 84 Look up table editor T...

Страница 35: ...In this example an input value of 5V will result in a reading of 10 and an input value of 5V will result in a reading of 50 Any input values between the points in this table will be linearly interpola...

Страница 36: ...formation is available together with the data from the transducer Smart Lemo connectors for voltages PRT s and Thermocouples are available from Chell To configure a channel for use with a Smart Lemo s...

Страница 37: ...e correct channel LUT config is in view in the Lookup Table Config window and add remove entry pairs manually or clear import an LUT from a file in the same way as with standard custom LUTs see sectio...

Страница 38: ...Page 37 5 9 Advanced Page The advanced tab contains extra options that users may find useful for more exact configuration but are not compulsory Figure 5 91 Advanced group...

Страница 39: ...nning average filter The impulse filter is a three term median filter designed to remove any noise on the data The average filter is a running average filter Select the appropriate filtering the click...

Страница 40: ...ps on the data packets If the timestamp is enabled it will have an effect on the maximum transmission rate Figure 5 95 Timestamp PTP synchronisation on checkbox This allows the user to select whether...

Страница 41: ...ps if the user is not using PTP In the first box it will show the timestamp and in the second box it displays the timestamp converted date time to make it easier to understand NOTE the user has to cli...

Страница 42: ...f the hardware trigger setting is requested as part of the c 00 command the FlightDAQ TL does not support a hardware trigger Transducer Heater Enable This is the function that allows the FlightDaq TL...

Страница 43: ...around 2Hz Figure 5 97 Live health check group 5 9 6 Live Data Precision This section in the Miscellaneous tab allows the user to configure the precision of the measurements displayed in the Live Dat...

Страница 44: ...es designed for the NetScanner products first enable the NetScanner emulation mode in the Advanced tab Click to enable NetScanner emulation and the enter the correct model type 9022 or 9146 Click Appl...

Страница 45: ...p expecting the default run state which is 0 This will expect secondary and tertiary data so you therefore get the following error when the Run screen is selected To overcome this another scan list mu...

Страница 46: ...Page 45 Select an additional Scan List in this case we have chosen 4...

Страница 47: ...Page 46 Then disable the secondary and tertiary data streams...

Страница 48: ...can the select Run first right click on the FlightDaq TL Acknowledge the two errors and then change the run state using the window below You will then be able to display the data from the FlightDAQ T...

Страница 49: ...evice to avoid damaging the gasket 3 Adjust DIP switch to required settings See table below 4 Change the metal identifying tag on the side of the case to show the correct mode DC or PoE 5 Re insert th...

Страница 50: ...g A dirty instrument may be wiped clean with a soft cloth that has been sprayed with a proprietary foaming cleaner then wiped dry immediately Under no circumstances should the instrument be wetted dir...

Страница 51: ...8mV 5V 10k 20 14 Thermocouple K 78mV 5V 10k 21 15 Thermocouple N 78mV 5V 10k 22 16 Thermocouple R 78mV 5V 10k 23 17 Thermocouple S 78mV 5V 10k 24 18 Thermocouple T 78mV 5V 10k 25 30 Resistance 250R N...

Страница 52: ...10V 10V 45 112 Pressure Ratiometric 78mV 10V 78mV 10V 10V 46 113 Pressure Ratiometric 300mV 10V 300mV 10V 10V 47 114 Pressure Ratiometric 5V 10V 5V 10V 10V 48 115 Pressure Ratiometric 10V 10V 10V 10V...