Page 29
5.8 Look-Up-Table Configuration
The FlightDaq-TL has a number of standard, in-built look-up-tables (LUT’s) for
converting the measured value to the appropriate engineering units.
In addition to this, the user is able to input a LUT for the primary and secondary
input of each individual channel and this is retained within memory in the
FlightDaq-TL. These LUT’s can have between 2 and 448 data pairs and the
FlightDaq-TL will linearly interpolate between the two closest values.
These LUT’s can also be saved in an appropriate Smart Lemo® connected
which enables the calibration for that transducer to be tied to, and travel with
that transducer.
5.8.1 Standard In-Built Look Up Tables
The standard in-built look-up-tables available are:
Transducer Type Transducer
variant
LUT Source Notes
RTD385
RTD385 / Pt100
Thermistor
10K
Thermocouple
B type
NIST
For all thermocouples, the Smart
Lemo® connector is used for the
cold junction. The cold junction
temperature is first converted into
a
voltage
(by
using
the
appropriate table in reverse)
which is then added to the
measured hot junction voltage.
This voltage is then converted into
temperature
by
using
the
appropriate table.
E type
NIST
J type
NIST
K type
NIST
N type
NIST
R type
NIST
S type
NIST
T type
NIST
Содержание FlightDaq-TL
Страница 20: ...Page 19 5 2 The Setup tab Figure 5 21 Main Setup page...
Страница 25: ...Page 24 Figure 5 52 Live Data Page Figure 5 53 Live Data Page...
Страница 46: ...Page 45 Select an additional Scan List in this case we have chosen 4...
Страница 47: ...Page 46 Then disable the secondary and tertiary data streams...