background image

1

User’s Guide

OMEGA® Scanner 3D

Содержание Scanner 3D

Страница 1: ...1 User s Guide OMEGA Scanner 3D...

Страница 2: ...mechanical for any purpose without the expressed written consent of The Ohio Willow Wood Company Copyright 2014 The Ohio Willow Wood Company All rights reserved OMEGA is a registered trademark of The...

Страница 3: ...3 TABLE OF CONTENTS Introduction 4 System Requirements 5 Hardware 6 Scanner Components 7 Setup 8 Calibration 9 Scanning 11 Maintenance 13 Technical Specifications 14...

Страница 4: ...e of the surface When the surface being scanned does not cause sufficient distortion of the light pattern reflective targets can be applied to the surface to provide additional detail WASTE ELECTRICAL...

Страница 5: ...ECOMMENDED SYSTEM REQUIREMENTS Operating System Windows 7 64 bit CPU Core i7 4 Cores Graphics NVIDIA Quadro K1000m or greater Memory 8 GB Hard Drive 320 GB Expansion Slot 2 0 USB port Display 1920 x 1...

Страница 6: ...ssary hardware fit into a protective case equipped with a custom foam insert to ensure that every part has its place The calibration plate is located under the foam insert Power Supply OMEGA Scanner 3...

Страница 7: ...scanner or any of its components Unauthorized opening of the scanner or its components could cause inadvertent damage and will void the warranty Please contact WillowWood in the event of any trouble...

Страница 8: ...SB cable before disconnecting anything else 1 Plug the power supply in to a wall outlet or power bar Connect the other end to the USB cable 2 Plug the USB cable into one of the USB 2 0 ports on the co...

Страница 9: ...ncluded with the Scanner The calibration plate should be at a temperature close to factory calbration conditions 20 22 C or 68 72 F The reflective targets on the calibration plate do not have the same...

Страница 10: ...ther 9 positions If the calibration procedure fails check the following If there are any reflective objects near the calibration plate remove the objects and repeat the calibration process Verify that...

Страница 11: ...you are prompted by the software to scan the model Hold the Scanner approximately 15 from the patient s body Aim the Scanner at an area of the surface that has the greatest amount of topographical de...

Страница 12: ...ce simply by flexing and extending your wrist as if you were painting with a brush will not be effective Avoid scanning any parts of the body that are moving for example if a child s shoulders are mov...

Страница 13: ...uire very special attention to ensure optimal quality 1 Spray dry air to remove any particles that may scratch the surface of the lenses and lights 2 Clean the objectives and lights using a damp cloth...

Страница 14: ...Resolution 1 mm 0 04 in Accuracy Up to 0 500 mm 0 02 in Volumetric Accuracy 0 500 mm m 0 02 in ft Stand Off Distance 400 mm 15 75 in Depth of Field from stand off distance 250 mm 10 in Scanning Area...

Страница 15: ...Darby Road Mt Sterling Ohio 43143 USA Phone 740 869 3377 or 800 848 4930 Fax 740 869 4374 www willowwoodco com Ohio Willow Wood Company B V Keizersgracht 62 64 1015 CS Amsterdam The Netherlands PN 232...

Отзывы: