Modifying Calibration Procedures
5.
Modify the Calibration Name. Press the
Enter
key to accept and automatically
advance to the next question in the calibration definition.
6.
Enter the background measurement time in minutes. Press the
Enter
key to
accept and automatically advance to the next question in the calibration
definition.
7.
Enter the number of individual background measurements. Press the
Enter
key to accept and automatically advance to the next question in the calibration
definition.
8.
Select the background blank source. Press the
Enter
key to accept and
automatically advance to the next question in the calibration definition.
9.
Enter the carrier number that contains the background blank for this
calibration. Press the
Enter
key to accept and automatically advance to the
next question in the calibration definition.
10.
Enter the
α
efficiency measurement time in minutes. Press the
Enter
key to
accept and automatically advance to the next question in the calibration
definition.
11.
Enter the number of individual
α
efficiency measurements. Press the
Enter
key to accept and automatically advance to the next question in the calibration
definition.
12.
Select the a efficiency source. Press the
Enter
key to accept and automatically
advance to the next question in the calibration definition.
13.
Enter the carrier number that contains the
α
efficiency source to be used for
this calibration. Press the
Enter
key to accept and automatically advance to
the next question in the calibration definition.
14.
Enter the
β
efficiency measurement time in minutes. Press the
Enter
key to
accept and automatically advance to the next question in the calibration
definition.
15.
Enter the number of individual
β
efficiency measurements. Press the
Enter
key to accept and automatically advance to the next question in the calibration
definition.
16.
Select the
β
efficiency source. Press the
Enter
key to accept and automatically
advance to the next question in the calibration definition.
17.
Enter the carrier number that contains the
β
efficiency source to be used for
this calibration. Press the
Enter
key to accept and automatically advance to
the next question in the calibration definition.
iMatic Automatic Counting System User's Manual - 9239221K
87