Define QC Procedures
To edit the Auto Cal Sequence Start:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “9” in the menu, Define
Calibrations, and then press the
Enter
key.
3.
Using the arrow keys, select the calibration (1 through 10), and then press the
Enter
key.
4.
Using the arrow keys position the cursor “>” to item “5”, AutoCal Sequence
Start, and then press the
Enter
key.
5.
Using the arrow keys to select the background or
α
and
β
efficiencies as the
first calibration procedure. Choose “4” if the background should not be
followed by efficiencies. If the calibration should not be automatically
executed at this time, select “Off”.
6.
Completing the Auto Cal Sequence Start menu will bring up a "Save
Changes?" question. Select
1. No
if the change was an error or
2. Yes
to
save the change(s) made.
To execute all defined calibrations make sure that the “CAL” group plate is in the
carrier stack ahead of the carriers defined for all active calibrations. Press
Count
to
execute calibrations.
Define QC Procedures
The Define QC Procedures menu is used to define and execute the daily Quality
Control procedures.
To access the Define QC Procedures menu:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “10” in the menu Define
QC Procedures, and press the
Enter
key.
10 Define QC Procedures
>
1 Select QC Reference
2 Daily Bkg Params
3 Daily a Eff Params
4 Daily b Eff Params
5 AutoCal Sequence Star
t
iMatic Automatic Counting System User's Manual - 9239221K
61