System Information
View DAQ BD Parameters
The View DAQ BD Parameters menu items allows you to view critical operating
parameters.
To view DAQ board parameters:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “4” in the menu, System
Information, and then press the
Enter
key.
3.
Using the arrow keys, position the cursor “>” to item “5”, View DAQ BD
Parameters, and then press the
Enter
key.
4.
Three items are displayed on this screen:
DAQBD Temp: 30.55
°
C (an example of DAQ board temperature)
Det Bias: 75.06 V (an example of actual detector bias)
Det Leakage: 3.45E-02
µ
A (an example of PIPS radiation detector leakage
current)
5.
These are read only fields. Pressing the
Enter
key will automatically advance
to the next item in the System Information list. Pressing
Cancel
will return to
the System Information menu. Pressing
Cancel
a second time will return you
to the Setup menu.
Version Information
The Version Information menu item allows you to view the version and date of the
firmware of the iMatic.
To display the version information:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “4” in the menu, System
Information, and then press the
Enter
key.
3.
Using the arrow keys, position the cursor “>” to item “6”, Version
Information, and then press the
Enter
key.
4.
This field is read only. Pressing the
Cancel
key will return to the System
Information Menu. Pressing the
Enter
key or pressing the
Cancel
key a
second time will return you to the Setup menu.
iMatic Automatic Counting System User's Manual - 9239221K
33