AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 251 of 402
Query Format
[SOURce:]CURRent:LIMit#:MAXimum?
Description
This command returns the maximum current limit setting available
for the referenced phase.
Returned Data Format
<nr2>
Query Example
CURR:LIM:MAX1?
41.67
Query Format
[SOURce:]CURRent:LIMit#:MINimum?
Description
This command returns the minimum current limit setting available for
the referenced phase.
Returned Data Format
<nr2>
Query Example
CURR:LIM:MIN1?
0.000
Command Syntax
[SOURce:]CURRent:LIMit:AUTO
Description
This commands enables or disables the Auto Current Limit function.
This function automatically adjusts the programmed current limit
level as a function of programmed voltage along the constant power
curve of the voltage range.
For example: On a 3150AFX model in three phase mode, if VOLT = 0
and CURR:LIM = 41.67 and voltage is changed to V = 300, the
CURR:LIM will change to: 5000 VA / 300 Vac = 16.0.
If CURR:LIM:AUTO = OFF then the CURR:LIM setting will remain at
41.67A and the power source will power limit at 5000VA instead if
the load draws more than 16Aac.
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<b>
Example
CURR:LIM:AUTO ON
Query Format
[SOURce:]CURRent:LIMit:AUTO?
Returned Data Format
<b>
Query Example
CURR:LIM:AUTO?
1