AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 204 of 402
Command Syntax
MEASure:LOGger:FILEName <cr>
Description
Sets the data logging file name
Parameters
Filename between double quotes
Parameter Format
<cr>
Example
MEAS:LOG:FILEN “Measurement-AC+DC”
Query Format
MEASure:LOGger:FILEName?
Returned Data Format
<cr>
Query Example
MEAS:LOG:FILEN?
Measurement-AC+DC
Command Syntax
MEASure:LOGger:MEMory <cr>
Description
Select the memory where the logging files will be saved. By default it
is RAM unless a memory stick or USB drive was inserted and selected
with this command:
MEAS:LOG:MEM:CAT? can be used to get the available memories.
Parameters
Memory name between double quotes.
Parameter Format
<cr>
Example
MEAS:LOG:MEM “USBA”
Query Format
MEASure:LOGger:MEMory?
Returned Data Format
<cr>
Query Example
MEAS:LOG:MEM?
USBA
Query Format
MEASure:LOGger:MEMory:CATalog?
Description
Returns the catalog of available memory names.
Returned Data Format
<cr>
Query Example
MEAS:LOG:MEM:CAT?
RAM, USBA
Command Syntax
MEASure:LOGger:RATE
Description
Sets the frequency at which the logging process will take
measurements.
Parameters
Frequency in Hz. Range is 1 Hz through 10 Hz.
Parameter Format
<nr1>
Example
MEAS:LOG:RATE 5
Query Format
MEASure:LOGger:RATE?
Returned Data Format
<nr1>
Query Example
MEAS:LOG:RATE?
5