Error Source
Code
Category
Error Message
Possible Cause
Resolution
Front panel
Error
130
Error
Unable to disable output.
One or more inverter
stages failed to disable.
Output OFF failed as one
of more inverters could
not be disabled
Turn Unit off with Circuit
Breaker
Front panel
Error
144
Error
Internal battery needs to
be replaced.
Warning. Front panel
controller battery low.
Front panel
Error
148
Error
Serial number mismatch:
Front panel does not
match power stages.
Setting conflict. Front
panel does not match
power stages serial
number
Contact customer service
Front panel
Error
150
Error
Suffix not allowed in
single form.
Setting conflict.
Command suffix invalid
for Single Phase mode
Change phase mode
Front panel
Error
151
Error
Suffix not allowed in split
form.
Setting conflict.
Command suffix invalid
for Split Phase mode
Change phase mode
Front panel
Error
152
Error
Peak current limit setting
is too low for the desired
voltage and frequency.
Setting conflict. Peak
current limit setting too
low to support requested
voltage and frequency.
AFX has a minimum peak
limit because it affects
output inductor current
and at high frequencies, a
circulating current
proportional to the
voltage and frequency
occurs.
Increase the peak current
limit setting
Front panel
Error
153
Error
Program transient trigger
input is disabled.
Setting conflict. External
Trigger Input disabled
Enable External Trigger
Input
Front panel
Error
154
Error
Steady state not yet
stabilized.
??
??
Front panel
Error
155
Error
Trigger output not
available in this model.
Trigger Output not
supported. Feature not
available
Requires AUX I/O board
Front panel
Error
156
Error
Trigger input not
available in this model.
Trigger Input not
supported. Feature not
available
Requires AUX I/O board
Front panel
Error
157
Error
Filter size must be an odd
number.
Invalid Filter Number.
Filter size must be odd
number
Change filter parameter
to even number
Front panel
Error
158
Error
Voltage mode not
available in UPC
compatible mode.
Setting conflict. UPC
mode only supports AC
Mode
Change to AFX Mode
Front panel
Error
159
Error
Waveform is unique in
UPC compatible mode.
Setting conflict.
Requested waveform
supported in UPC mode
only
Change to UPC Mode
Front panel
Error
160
Error
Phase is 180 for split in
UPC compatible mode.
Setting conflict. Phase
angle 180 reserved for
Split phase mode in UPC
Mode
Change to AFX Mode
Front panel
Error
161
Error
Voltage AC is unique in
UPC compatible mode.
Setting conflict. UPC
mode only supports AC
Mode
Change to AFX Mode