No
Yes
Lan Save
Select whether to save the LAN setting
or not.
No
Yes
RS232
+CAN
When insert IT-E1207 communication board into expansion
slot, the menu displays this information.
RS232
RS232 communication interface.
Display as the following format: baud rate_data bit_
parity bit_stop bit.
•
Select the baud rate from the following options:
4800, 9600, 19200, 38400, 57600, 115200
•
The data bit, parity bit and stop bit are fixed: 8_
None_1
CAN
CAN communication interface.
Select the baud rate from the following options: 20k,
40k, 50k, 80k, 100k, 125k, 150k, 200k, 250k, 400k,
500k, 1000k.
Address: Set the instrument address to a number
from 1 to 127.
RS485
+Ext
When insert IT-E1208 communication board into expansion
slot, the menu displays this information.
Analog
Set external analog interface state.
RS485
RS485 communication interface.
Display as the following format: baud rate_data bit_
parity bit_stop bit.
•
Select the baud rate from the following options:
4800, 9600, 19200, 38400, 57600.
•
The data bit, parity bit and stop bit are fixed: 8_
None_1.
•
RS485 Addr: Set the instrument address to a
number from 1 to 127.
GPIB
When insert IT-E1205 communication board into expansion
slot, the menu displays this information.
Address: Set the instrument address to a number from 0 to 30.
USB
When insert IT-E1209 communication board into expansion
slot, the menu displays this information.
Parallel
Set the instruments to parallel operation mode.
Single
Set the instrument to single mode.
Slave
Set the instrument to slave mode.