Capacity: Battery capacity
—EXIT—: Return to the previous level
Parallel
Set the number of emulation batteries in
parallel connection
Series
Set the number of emulation batteries in
serial connection
I
Positive current limit value, maximum
discharge current of battery emulation
group
I_Limit –
Negative current limit value, maximum
charge current of battery emulation
group
—EXIT—
Return to the previous level
Edit
Edit battery emulation file
Edit File=1/10
Select the number of file to be edited. A maximum of 10
List files can be edited.
Cell
Properties
Battery attribute setting, to be set based
on the emulation battery specification
Full Volt: Battery voltage value under full
state
Empty Volt: Battery voltage value under
empty state
Inner R: Battery internal resistance
value
Capacity: Battery capacity
—EXIT—: Return to the previous level
Parallel
Set the number of emulation batteries in
parallel connection, with setting range
of: 1–99
Series
Set the number of emulation batteries in
serial connection, with setting range of:
1–99
I
Positive current limit value, maximum
discharge current of battery emulation
group
I_Limit –
Negative current limit value, maximum
charge current of battery emulation
group
—EXIT—
Return to the previous level
Operation steps
1. Under the source mode, press
[Shift]+[I-set]
(Function) composite key, en-
ter the Function menu page.
Copyright © Itech Electronic Co., Ltd.