7.4.2 Set the Power-on State
This menu item is used to control the settings and output state when the power
system is powered on.The detailed power-on parameters and output state are
as follows.
•
The run mode of instrument and setting value.
•
The output state of the power system, i.e., the
[On/Off]
key state.
Operation procedures:
1. Press the composite keys
[Shift]+[P-Set]
(System) to enter System menu
interface.
2. Rotate the knob to select
PowerOn
and press
[Enter]
to confirm.
3. Rotate the knob to set the power on state and press
[Enter]
to confirm.
•
Reset: default value, indicates when the instrument is powered on, the
instrument will initialize the *RST setting.
•
Last: indicates when powered on, the instrument will remain the same
settings and output state as last time you turned off.
•
Last+Off: indicates when powered on, the instrument will remain the
same settings as last time you turned off, but the output state is OFF
state.
4. Press
[Esc]
to exit menu interface.
7.4.3 Set the Sense Function State
This menu item is used to switch the instrument to local sensing or remote
sensing. This series instrument support two connection methods: Local sensing
and Remote sensing. The remote sensing is used for maximizing measurement
accuracy. (Refer to
2.5 Connecting the Device Under Test (DUT)
for more
information). The procedures to set the menu item are as follows.
1. Press the composite kyes
[Shift]+[P-Set]
(System) to enter the System
menu interface.
2. Rotate the knob to select
Sense
and press
[Enter]
to confirm.
3. Rotate the knob to select
On
or
Off
and press
[Enter]
to confirm.
•
Off: default value, indicates turn the sense function off.
•
On: indicates turn the sense function on.
4. Press the
[Esc]
key to exit.
7.4.4 Set the Communication Information
This menu item is used to set the communication information between
instrument and PC. There is no standard communication interfaces for this
Copyright © Itech Electronic Co., Ltd.