2. Select the desired file number to be edit. The maximum file number is 10.
Edit File = 1/10
3. Press
[Enter]
to confirm. Take the charging mode as an example.
1
:
Batt Mode= Charge
2
:
Charge V = 1.00V
3
:
Charge I = 1.00A
4
:
Charge Time = 60S
5
:
Cut Off V = 8.00V
6
:
Cut Off I = 5.00A
7: Cut Off Q = 10AH
When Batt Mode set to Discharging, perform discharging test, discharging
voltage/current/time can be displayed in the user interface. Users according
to needs to choose discharging or charging mode.
4. Set the attribute parameters of the List file in successive. After setting press
[Esc]
to exit.
Recall and view battery test file
1. In the battery test function setting interface, select the
Recall
and enter to
the edit file menu.
2. Rotate the knob to set the number of List file to be recalled. A maximum of
10 battery test files can be recalled in this instrument.
Recall Batt = 1/10
When the selected number is empty, the interface would prompt that the
recalled file is empty. The user needs to select the file again or return and
enter the Edit menu for editing.
3. After successful recall, the user can view the setting information of the
existing battery test file. The test parameters can only be viewed, and
cannot be edited. To modify them, please return and enter the Edit menu to
select the corresponding file number for editing.
4. Press
[Esc]
key, and exit the existing page.
Run Battery test file
After the battery test files are edited, the user needs to select
Run
to enable the
battery test function and return to the main interface to trigger the running.
12.000V
10.000A
00:00:01
0.00Ah
•
First line: The voltage and current value.
•
Second line: testing time and battery capacity.
Copyright © Itech Electronic Co., Ltd.