Visual-ICE™ Cryoablation System
7-9
7.4
Mechanical Issues
Symptom
Potential Causes / Solutions
Needle cannot be firmly connected
in the needle port
1. Verify that the locking bar is in the UNLOCKED position.
2.
Needle connector may be defective. Use a different
needle.
3. Residual gas pressure may be in the needle port. Use a
different channel.
4. Check the
Gas Indicator
display. If there is pressure in
the system, end the procedure and vent the system using
the automatic vent function.
Locking bar on the Needle
Connection Panel cannot be
positioned in the LOCKED position
1. Verify that all needles in the channel are fully inserted in
the needle connection ports.
2. The Locking Bar may be defective. Move the needle to a
different channel. Contact Galil Medical customer service
to schedule a service call.
3. Check the Gas Indicator display and verify the system is
not pressurized. If there is pressure in the system, end
the procedure and vent the system using the automatic
vent function.
Console does not move freely
1. Release the brake to unlock the front wheels.
2. Check the individual rear wheel brakes and ensure the
brakes are released.
7.5 Gas Cylinders and Gas Supply Lines
Symptom
Potential Causes / Solutions
Safety cable is missing at either the
cylinder side or the system side of
the gas supply line
Do not use a gas supply line with a missing safety cable.
Doing so could compromise the safety of personnel in the
room. Contact Galil Medical Customer Service.
A gauge or a gas supply line is
damaged
Do not use any damaged product. Contact Galil Medical
Customer Service for new accessories.
A gas leak was detected between
the gauge adapter and the cylinder
main valve
•
Tighten the connection using the wrench provided with
the Visual-ICE System.
•
Close the cylinder valve and vent the gas from the
Visual-ICE System and the gas supply lines using the
Manual Vent Valve (Figure 4-1). Verify the system is
depressurized. Loosen and remove the gauge assembly
adapter. Verify no debris is on the gas cylinder connection
point; clean the sealing surface as needed to remove
any debris. Reposition and tighten the gauge assembly
adapter onto the cylinder valve, using the wrench
provided with the Visual-ICE System.
Содержание Visual-ICE
Страница 1: ...User Manual ...
Страница 10: ...viii User Manual This page intentionally left blank ...
Страница 22: ...1 10 User Manual This page intentionally left blank ...
Страница 30: ...2 8 User Manual This page intentionally left blank ...
Страница 46: ...3 16 User Manual Screen 3 16 Remote Upload Download Screen Screen 3 17 Successful Upload Download ...
Страница 76: ...4 30 User Manual This page intentionally left blank ...
Страница 114: ...7 32 User Manual This page intentionally left blank ...
Страница 122: ...9 2 User Manual This page intentionally left blank ...
Страница 124: ...10 2 User Manual This page intentionally left blank ...
Страница 125: ......