Visual-ICE™ Cryoablation System
4-23
Screen 4-21.
Cautery Confirmation message
3.
Press the desired needle port (A or B) on each channel containing a needle you wish to track
ablate. Only one needle per channel may activate the cautery function at a given time.
NOTE:
The Visual-ICE System uses a pre-set duration of 30 seconds for each cautery phase for
1.5 CX-type needles. This value is not adjustable, but the duration can be stopped in advance of
30 seconds by pressing the
Stop
button.
NOTE:
A maximum of four (4) needles may use the cautery function simultaneously.
4. Press the
Start
button to initiate track ablation (Cautery).
•
During the needle warming phase, the Cautery Control Screen will display a rotating warming
Screen 4-22.
1.5 CX Needle Warming
•
When the cautery threshold temperature is reached, Cauterization automatically begins.
o
The timer will display Cauterizing (Screen 4-23).
o
The Cautery Control Screen will display the estimated temperature range for the needle
shaft.
NOTE:
The shaft temperature is displayed as a temperature range because tissue and
procedural variables affect the temperature.
Screen 4-23. 1.5 CX Needle Cautery In Progress
Содержание Visual-ICE
Страница 1: ...User Manual ...
Страница 10: ...viii User Manual This page intentionally left blank ...
Страница 22: ...1 10 User Manual This page intentionally left blank ...
Страница 30: ...2 8 User Manual This page intentionally left blank ...
Страница 46: ...3 16 User Manual Screen 3 16 Remote Upload Download Screen Screen 3 17 Successful Upload Download ...
Страница 76: ...4 30 User Manual This page intentionally left blank ...
Страница 114: ...7 32 User Manual This page intentionally left blank ...
Страница 122: ...9 2 User Manual This page intentionally left blank ...
Страница 124: ...10 2 User Manual This page intentionally left blank ...
Страница 125: ......