Visual-ICE™ Cryoablation System
4-25
guide needle removal. The Active Zone Indicator is a marked band along the needle shaft,
located 20 mm distal to the heating portion of the needle.
4.7 Advanced Channel Controls
Advanced Channel Controls for each channel provide options to change the needle type for the
selected channel, link two channels together, and program multiple freeze-thaw cycles.
4.7.1 Select Needle Type Control
1.
To change the needle type for a channel, press and hold the
Channel
button to enter Advanced
Channel Controls for that channel (Screen 4-24).
2.
Select the proper needle type from the drop-down menu.
3. Press the
OK
button.
Screen 4-24. Advanced Channel Controls
4.7.2 Link Channels Control
1. Press and hold the
Channel
button to enter
Advanced Channel Controls
for that channel
2.
Press the
Link
button to link two channels together for simultaneous operation. When two
channels are linked, the
Channel
button shows both channels (Screen 4-25).
NOTE
: This function is not available on the channel labeled
ALL
. You can only link channels
that are on the same horizontal plane on the Needle Connection Panel (e.g., 1 and 2, 3 and 4,
5 and 6).
Screen 4-25. Linked Channels
Содержание Visual-ICE
Страница 1: ...User Manual ...
Страница 10: ...viii User Manual This page intentionally left blank ...
Страница 22: ...1 10 User Manual This page intentionally left blank ...
Страница 30: ...2 8 User Manual This page intentionally left blank ...
Страница 46: ...3 16 User Manual Screen 3 16 Remote Upload Download Screen Screen 3 17 Successful Upload Download ...
Страница 76: ...4 30 User Manual This page intentionally left blank ...
Страница 114: ...7 32 User Manual This page intentionally left blank ...
Страница 122: ...9 2 User Manual This page intentionally left blank ...
Страница 124: ...10 2 User Manual This page intentionally left blank ...
Страница 125: ......