Visual-ICE™ Cryoablation System
7-21
Message
Reason for Appearance
40-08 i-Thaw Error
•
Channel [Xa or Xb] i-Thaw needle is defective for
electrical thawing
•
Use passive thaw or connect helium gas
An i-Thaw needle was identified as not
i-Thaw operational. Passive thaw or
connection of helium for active thaw is
required.
TEMPERATURE SENSOR / ORGAN MAP
50-01 MTS Alarm
•
Sensor point (5, 15, 25 or 35) on MTS (#X) met
selected alarm limit
The MTS detected the selected alarm
temperature.
50-02 MTS Alarm
•
Sensor point (5, 15, 25 or 35) on MTS (#X) met
selected alarm limit
The MTS detected the selected alarm
rate of temperature change within 30
seconds.
50-03 MTS Disconnected
•
An MTS is disconnected
•
Reconnect the MTS to continue to monitor
temperature in that location
An MTS became disconnected during a
procedure.