Visual-ICE™ Cryoablation System
4-21
•
Similarly, when
FastThaw
is selected, pressing the
Change to i-Thaw
button will change the
thaw from FastThaw to i-Thaw (Screen 4-18).
Screen 4-18. Advanced Thaw Controls i-Thaw
2. After selecting the type of thaw, press the
Thaw or FastThaw
button in the
Channel Control
to
initiate thawing.
NOTE:
The FastThaw function generates a temperature that is higher than that generated for
i-Thaw, resulting in a faster thaw time.
NOTE:
A maximum of 7 needles can thaw simultaneously using the i-Thaw function; a maximum
of 4 needles can thaw simultaneously using the FastThaw function.
3.
During the needle warming phase for CX-type needles, the channel status will display a rotating
warming indicator (Screen 4-19).
Screen 4-19. Advanced Thaw - Needle Warming
4.
When the threshold temperature for i-Thaw or FastThaw is reached, the channel status will
display the estimated temperature range for the needle shaft (Screen 4-20).
NOTE:
The shaft temperature is displayed as a temperature range because tissue and
procedural variables will affect the temperature.
Содержание Visual-ICE
Страница 1: ...User Manual ...
Страница 10: ...viii User Manual This page intentionally left blank ...
Страница 22: ...1 10 User Manual This page intentionally left blank ...
Страница 30: ...2 8 User Manual This page intentionally left blank ...
Страница 46: ...3 16 User Manual Screen 3 16 Remote Upload Download Screen Screen 3 17 Successful Upload Download ...
Страница 76: ...4 30 User Manual This page intentionally left blank ...
Страница 114: ...7 32 User Manual This page intentionally left blank ...
Страница 122: ...9 2 User Manual This page intentionally left blank ...
Страница 124: ...10 2 User Manual This page intentionally left blank ...
Страница 125: ......