7-30
User Manual
Message
Reason for Appearance
80-34 System Error
•
Fan X failure. At the conclusion of the procedure,
contact Galil Medical Customer Service.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
Fan X was identified as not operating.
80-35 System Error
•
The gas flow rate in channel XX exceeds the
system operation requirements and may impact
performance. Reduce the number of active needles.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
The calculated flow rate for a given
channel exceeded the optimum system
requirement. The number of active
needles should be reduced.
80-36 System Error
•
The gas flow rate exceeds the system operation
requirements and may impact performance. Reduce
the number of active needles
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
The cumulative flow rate for all channels
exceeded the optimum system
requirement. The number of active
needles should be reduced.
Содержание Visual-ICE
Страница 1: ...User Manual ...
Страница 10: ...viii User Manual This page intentionally left blank ...
Страница 22: ...1 10 User Manual This page intentionally left blank ...
Страница 30: ...2 8 User Manual This page intentionally left blank ...
Страница 46: ...3 16 User Manual Screen 3 16 Remote Upload Download Screen Screen 3 17 Successful Upload Download ...
Страница 76: ...4 30 User Manual This page intentionally left blank ...
Страница 114: ...7 32 User Manual This page intentionally left blank ...
Страница 122: ...9 2 User Manual This page intentionally left blank ...
Страница 124: ...10 2 User Manual This page intentionally left blank ...
Страница 125: ......