Visual-ICE™ Cryoablation System
6-1
6 SYSTEM CARE and MAINTENANCE
6.1 System Installation
Installation instructions for the Visual-ICE™ System are included in the shipping carton, available
online at www.galilmedical.com/customerservice, or by contacting your local Galil Medical
representative.
6.2 Cleaning
6.2.1 Cleaning the Visual-ICE System
Clean the Visual-ICE Cryoablation System after each use by following the steps below.
1. Clean the Touch Screen Monitor when the Visual-ICE Cryoablation System is powered OFF.
•
Wipe the screen gently with a damp gauze wipe.
•
Use water or isopropyl alcohol only.
•
Do not use cleaning agents such as Betadine or bleach solution.
2. Clean the Visual-ICE Cryoablation Console by wiping with a damp gauze wipe.
•
Use soap and water or isopropyl alcohol cleaning solutions.
•
Do not use cleaning agents such as Betadine or bleach solution.
•
Do not allow water or other liquid to drip or seep inside the cryoablation needle connection
ports. The needle connection ports must remain completely dry at all times.
3. Ensure that cleaned surfaces are dry prior to closing or turning on the system.
CAUTION.
Do not set food, drinks, or other objects on top of the Visual-ICE Cryoablation System.
Doing so may damage the system.
6.2.2 Cleaning the Stylus
Clean the stylus after each use by wiping down with isopropyl alcohol. Dry the stylus with a dry
gauze wipe.
6.3 Service and Preventive Maintenance
Only trained and authorized service engineers are permitted to conduct service and system
preventive maintenance on the system. Preventive maintenance on the Visual-ICE Cryoablation
System will be performed every two years.
CAUTION.
Do not attempt to service the Visual-ICE Cryoablation System. If service is required,
contact Galil Medical Customer Service.
The Visual-ICE Cryoablation System will display a reminder on the screen approximately one
month prior to when the preventive maintenance is due. If the reminder message appears and the
preventive maintenance has not yet been scheduled, contact your Galil Medical authorized service
representative to schedule the service.
Содержание Visual-ICE
Страница 1: ...User Manual ...
Страница 10: ...viii User Manual This page intentionally left blank ...
Страница 22: ...1 10 User Manual This page intentionally left blank ...
Страница 30: ...2 8 User Manual This page intentionally left blank ...
Страница 46: ...3 16 User Manual Screen 3 16 Remote Upload Download Screen Screen 3 17 Successful Upload Download ...
Страница 76: ...4 30 User Manual This page intentionally left blank ...
Страница 114: ...7 32 User Manual This page intentionally left blank ...
Страница 122: ...9 2 User Manual This page intentionally left blank ...
Страница 124: ...10 2 User Manual This page intentionally left blank ...
Страница 125: ......