Visual-ICE™ Cryoablation System
7-29
Message
Reason for Appearance
80-31 System Error
•
Gas cylinder shut off valve is not open enough to
provide adequate flow. If necessary, open the valve
approximately another half turn.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
Gas flow from gas cylinder was
insufficient. To improve gas flow, the
cylinder valve should be opened further.
80-32 System Error
•
Channel X is defective. Choose another channel.
At the conclusion of the procedure, contact Galil
Medical Customer Service.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
A defective solenoid on Channel X was
detected; another channel should be
selected.
80-33 System Error
•
Defective MTS in channel [X]. Replace with a new
MTS.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
The MTS temperature display was out
of the expected range when initially
connected to the needle connection
panel.
Содержание Visual-ICE
Страница 1: ...User Manual ...
Страница 10: ...viii User Manual This page intentionally left blank ...
Страница 22: ...1 10 User Manual This page intentionally left blank ...
Страница 30: ...2 8 User Manual This page intentionally left blank ...
Страница 46: ...3 16 User Manual Screen 3 16 Remote Upload Download Screen Screen 3 17 Successful Upload Download ...
Страница 76: ...4 30 User Manual This page intentionally left blank ...
Страница 114: ...7 32 User Manual This page intentionally left blank ...
Страница 122: ...9 2 User Manual This page intentionally left blank ...
Страница 124: ...10 2 User Manual This page intentionally left blank ...
Страница 125: ......