4 Well Manifold Test Parameters
Evacuation Efficiency Test
Residual QC Test
Aspirate Height: Z Offset
13 steps
1.65 mm
Y Offset Position
-90
4.11 mm
Aspirate Travel Rate
1
5.0 mm/sec
Aspirate Delay
2000 msec
Secondary Aspirate
No
Dispense Precision Test
Accuracy QC Test
Dispense Volume
1120 µL/well
Dispense Flow Rate
5
Dispense Height: Z Offset
180 steps (22.87 mm)
Y Offset
00
Prime Before Start
Yes
Prime volume and flow rate
5 mL at rate 5