![Biotek 50 TS Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/biotek/50-ts/50-ts_instructions-for-use-manual_2756089040.webp)
| 11
BioTek Instruments, Inc.
Protocol Name
COSTAR_
FLT16_96
COSTAR_
FLT16_384
NUNC_
FLT16_96
NUNC_
FLT16_384
Final Aspirate
Yes
Yes
Yes
Yes
Final Aspirate Delay
0
0
0
0
Biometric Separation Sample Protocols for 50TS: 8M, 8MV, 8MF Models
Protocol Name
LUM_MAG_FLAT_96 LUM_MAG_RND_96
Plate Type
96
96
Shake 1 sec, 1 (15Hz), soak 1 min
Number of Cycles
2
2
Wash Format
Plate
Plate
Soak After Dispense
Yes, 30 sec
Yes, 30 sec
Shake After Soak
No
No
Prime After Soak
No
No
Dispense Volume
200
100
Dispense Flow Rate
3
5
Dispense Height
120
128
Horizontal Dispense Position
0
0
Bottom Wash First
No
No
Prime Before Start
Yes
Yes
Prime Volume
5
5
Prime Flow Rate
5
5
Aspirate Height
29
58
Horizontal Aspirate Position
0
24
Aspirate Rate
7
1
Aspirate Delay
0
0
Crosswise Aspiration
No
No
Final Aspirate
Yes
Yes
Final Aspirate Height
38
58
Final Horizontal Aspirate Position 0
24
Final Aspirate Rate
7
1
Final Aspirate Delay
0
0