16 |
50™ TS Washer
3.
Clean the dispense and aspirate chambers on the manifold top with hot
soapy water and a soft-bristled brush. Thoroughly clean the walls of the
chambers.
4.
Clean the insides of the aspirate and dispense tubes with the appropriate
stylus (aspirate or dispense).
5.
Rinse the manifold with deionized or distilled water. Look through the
tubes to ensure tube openings are clear.
6.
When satisfied, replace the gasket and metal plate. Do not overtighten.
7.
Remount the manifold and replace the tubing.
8.
Replace the mist shield.
9.
Run a Prime program using 200 mL.
10.
Verify dispenser performance visually or by completing the
Dispense
Precision Test 8, 8s, 2x8, and 12 Manifolds
.
8,16 Well Manifold:
Important: Do not disassemble the 8,16 Well manifold. There are five screws
on the top metal plate. Removing any of these screws will cause the manifold to
fall out of alignment!
1.
Remove the manifold:
•
Disconnect the tubing from the back of the manifold.
•
Remove the three thumbscrews on top of the manifold.
•
Gently grasp the front of the manifold, holding the together the two rows of
tubes and lift the manifold straight up and off.