![Biotek 50 TS Скачать руководство пользователя страница 77](http://html.mh-extra.com/html/biotek/50-ts/50-ts_instructions-for-use-manual_2756089077.webp)
2 |
50™ TS Washer
About Instrument Qualification
Instrument qualification for the 50 TS involves three activities: qualifying
installation and setup, qualifying wash, dispense and aspirate routines, and
qualifying long-term stability. This is referred to as Installation Qualification
(
IQ
), Operational Qualification (
OQ
), and Performance Qualification (
PQ
),
respectively.
Testing includes:
•
The System Self Test verifies system components, such as the vacuum,
manifold, and carrier positioning.
•
Evacuation Efficiency: This test measures the residual volume per well after
the aspiration step of plate washing. The lower the residuals per well, the better
the evacuation efficiency of the washer.
•
The Evacuation Efficiency Test for Vacuum Filtration verifies that the
washer effectively aspirates fluid from the wells of a 96-well filter microplate.
The test measures the total residual volume of the plate after evacuation to
verify it is performing to specifications.
•
Dispense Precision: This test measures the variability of volumes dispensed
from tube to tube across the manifold.
•
Annual Buffer Switching Test: The Dispense Precision Test for the Buffer
Switching valves measures the variability of volumes dispensed from tube to
tube across the manifold when the washer is equipped with buffer-switching
valves.
Qualification Schedule
The following schedule defines the recommended intervals for verification
tests for a washer used two to five days a week. The schedule assumes that the
washer is properly maintained as outlined in
Recommended Maintenance
Schedule
.
Note: An instrument qualification package (PN1550510) for the
washer is available for purchase. The package contains thorough
procedures for performing Installation Qualification, Operational
Qualification and Performance Qualification (IQ-OQ-PQ) and
preventative maintenance (PM). Extensive Checklists and Logbooks are
included for recording results. Contact your local dealer or BioTek for
more information.