6 |
50™ TS Washer
Throughput (Processing Speed)
Testing is performed with the Dispense Rate 9 (Rate 5 for 2x8 Manifold),
Aspirate Rate 7 and using a Strip Wash Format. Timing is started when the
manifold is over the first strip and is stopped when the last strip has been
aspirated.
Manifold
Type
Plate
Type
Performance
8 and 8s Well 96-well
<130 seconds for 12 strips (3 cycles, 300 µL/well, no
soak)
8,16 Well
384-well <260 seconds for 24 strips (3 cycles, 100 µL/well, no
soak)
2x8 Well
96-well
<80 seconds for 12 strips (3 cycles, 300 µl/well, no soak)
4 Well
24-well
<60 seconds for a 24-well plate (1 cycle, 1120 µl/well, no
soak)