20 |
50™ TS Washer
Options
Actions
Default
Secondary
2
aspirate
(Crosswise)
A secondary or crosswise aspiration is a two-step
aspiration. The wells are first aspirated at the horizontal
aspirate position. The aspirate tubes rise a fixed number
of steps and then do a second aspiration at the secondary
position. It is useful for eliminating residual sample or
reagent from the wall perimeter.
No
Final Aspirate
Select YES to designate a final aspiration, leaving the
wells empty.
Yes
Final Aspirate
Delay
Same as Aspiration Delay (see above) applied to the final
aspiration.
0
•
Negative values: To enter a negative value, toggle the +/- button.
2
Secondary and Final Aspiration are only available in a wash step, not stand-alone aspiration
steps.
Soak and Shake Protocol Parameters
Add a Shake/Soak step to the protocol to mix plate contents.