| 23
BioTek Instruments, Inc.
Instrument Configuration - Setup Controls
The
Instrument
menu leads to controls for:
o
Instrument Setup below
o
AutoPrime
Instrument Settings
Manifold Selection
Your 50 TS was initially configured for and shipped with a wash
manifold. The washer’s manifold setting must match the type of manifold
installed on the instrument. If you change the hardware, you must also
change the setting.
Manifold Type
PN
Plate Type
Onboard
4 Well
1550504
24-well
4
8 Well
4070512
96-well
8
8 Well Short Dispense
Tube
4070519
96-well
8s
2 X 8 Well
1550501
96-well
2x8
12 Well
4070513
96-well
12
8, 16 Well
4070527
96 & 384-well
8,16
Plate Carrier Selection
The type of plate carrier installed on the instrument must match this setting to
ensure expected performance.
•
Standard
•
384 Well
: compatible with the 8,16 Well manifold.
•
Magnetic Separation
: special plate carrier for holding a magnet.
•
Vacuum Filtration
: special carrier for performing vacuum aspiration or
filtration assays. This setting must be selected to run a vacuum filtration
protocol.