BioTek Instruments, Inc.
Decontamination Procedure for Tubing and Manifold
Two supply bottles are required for this procedure: one for 120 mL
disinfectant, and one for 250 mL of rinse.
After decontaminating the external surfaces:
1.
Connect the washer to the power supply, and turn it on.
2.
Connect the disinfectant fluid supply and empty the waste bottle.
3.
Run
DECONTAMINATION
. (Select
Maintenance
from the Main Menu.)
While the program is running, you will need to periodically check the
display and follow instructions for changing the supply bottles.
4.
Flush all the bleach/disinfectant from the instrument by running
DAY_RINSE
at least once with DI water.