| 9
BioTek Instruments, Inc.
3.
Remove the waste bottle cover and replace it with the stopper from the
waste tubing set. Take note of the distinct fittings at the end of each tube:
one connects to the
Vacuum
port, the other to the
Waste
port on the back
of the washer.
4.
To install the optional inline vacuum filter:
1.
Cut the tubing that will connect the waste bottle to the
Vacuum
port
leaving about an inch of tubing extending from the washer,
2.
Insert the vacuum filter between the cut ends of the tubing.
5.
Attach the tubing from the waste bottle to their respective ports on the back
of the washer:
•
Vacuum
port to connect the vacuum pump to the waste bottle,
•
Waste
port to capture waste fluid from the washer.
6.
Connect the tubing from the vacuum filtration carrier to the
Carrier
port on
the back of the washer.
Disconnect the waste tube from the Carrier port when you are not
using the vacuum filtration carrier. An internal valve preserves the
vacuum required to perform standard aspiration.
Install the Liquid Level Alert System (Optional)
The Liquid Level Alert
select box
has ports, labeled A, B, and C, for the
supply bottles, and Waste.
1.
If you have not already done so, connect the waste and fluid supply
bottles as described on the previous pages.
2.
Connect the cables from the supply and waste bottles to the select box: