Alarm and Alarm Status
High P-wave and T-wave, Long P-wave or T-wave with high amplitude duration can be detected
by QRS Complex. Place the leads on the ECG1 channel with the highest R-wave (compare to T-
wave and / or P-wave) to allow the monitor to properly detect low heart rate conditions in this
situation. If the monitor continues to misinterpret the P-wave or T-wave, use a pulse oximeter to
reposition the electrodes or monitor the animal's pulse rate.
Display
①
②
③
④
⑤
①
Heart rate detector: It detects heart rate and flickers simultaneously.
②
Pace maker: Pace maker signal is detected and flashes simultaneously.
③
Heart rate: Displays the heart rate per minute.
④
HR Alarm limits: Heart rate upper and lower thresholds are displayed.
⑤
PVC count number per 1 minute is display.
ECG Settings
Main menu
Sub menu
ECG
A.
ECG Parameters
A-1.
Alarm
A-2
. QRS Volume
A-3.
Display Option
A-4.
Arrhythmia
A-5.
ST/PVC
A-6.
Pace Maker
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...