177
BM5Vet Pro
User’s Manual
20. Technical Specification
Overview
The monitor is not user installable. It must be installed by qualified service personnel.
The monitor is intended to be used for monitoring, recording, and alarming of multiple
physiological parameters of animals in health care facilities. The device is to be used by trained
health care professionals.
The monitor is intended for use in health care facilities; the BM5 Vet Pro Monitor is additionally
intended for use in transport situations within the hospital setting.
EMC Compatibility (EMC)
Much of the information below has been borrowed from the requirements set forth in
the Electromagnetic Compatibility Standard IEC 60601-1-2 for medical electrical
equipment issued by the International Electro technical Commission and is available from
a variety of sources. Although primarily aimed at equipment manufacturers, most of the
information contained here is useful for users interested in medical equipment.
The information contained in this section (such as separation distance) is generally
information about the Bionet Veterinary patient monitor detailed above. The numbers
provided here are not guaranteed but are provided with reasonable assurance of error-
free operation. This information may not apply to other medical and electrical systems,
and older equipment may be particularly susceptible to interference.
Note
Medical electrical equipment requires special precautions for electromagnetic compatibility and
must be installed and serviced in accordance with the EMC information in this section and in the
operating instructions supplied with the monitor.
Portable and mobile RF communication equipment can affect medical electrical equipment.
Cables and accessories not specified in the user guide are not certified. Using other cables and
/ or accessories may adversely affect safety, performance, and electromagnetic compatibility
(increased electromagnetic emissions and reduced immunity).
This equipment should not be used near or on top of other equipment. If you need to
use it on its side or stacked, you should observe the equipment to make sure it works
properly within your configuration.
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...