72
BM5Vet Pro
User’s Manual
A.
ECG menu
MENU
Description
Available settings
A-1.
Alarm
ECG alarm setting menu
A-1-1.
PARAMETER ALARM
LIMIT
HR, ST, PVC parameter alarm limits,
level, activation setup menu.
A-1-2.
TECHNICAL ALARM
CONDITION
ECG-CABLEOFF
ECG-LEADFAULT
ECG-CHECKELECTRODE
ECG-HR-SEARCH
A-2.
QRS VOLUME
QRS detection volume setting menu.
When you set the SpO2 volume, it is
automatically set to OFF.
OFF, 0%~100%
A-3.
DISPLAY OPTION
A-3-1.
SWEEP SPEED
The speed of the ECG displayed on the
screen can be set. Default setting:
25mm/s
6.25mm/s, 12.5mm/s,
25mm/s, 50mm/s
A-3-2.
FILTER
The filter setting is MONITOR by
default.
ECG FILTER
:
Selects among four
frequency bands to filter the signal.
MONITOR 0.5Hz ~ 40Hz
MODERATE 0.5Hz ~25Hz
MAXIMUM 5Hz ~ 25Hz
DIAGONOSIS 0.05Hz ~150Hz
MONITOR
MODERATE
MAXIMUM
DIAGONOSIS
A-3-3.
SIZE (SENSITIVITY)
Changes the display amplitude of the
ECG waveform.
0.25, 0.5, 1, 2,
4mm/mV
A-3-4.
HR SOURCE
The cardiac source can be selected as
ECG or SpO2, AUTO.
ECG, SpO2, AUTO
A-3-5.
VIEW CHANNEL
Set the number of channels of ECG
waveform to be displayed on the
screen.
1CH,
2CH,
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...