166
BM5Vet Pro
User’s Manual
18.
Maintenance And Troubleshooting
Noise in ECG
-
Make sure your filter settings are appropriate
-
Make sure the electrode is attached
-
Make sure the gel on the electrode is not dry.
The Patient
’s skin Yes
is extremely dry
The electrode is Yes
in poor condition
The cable(lead wire) Yes
Is disconnected
Apply ECG gel or water to the area of
contact then use the electrode
Replace the electrodes (Use CE
compatable products
)
Replace lead wires
Repair the ECG B/D