Data storage failure
"Admit" has been selected in the menu
The Digital B/D(Digi)
Battery has been lower than 3V
No
No
Execute the "admit" function
(No data will be stored during battery discharge)
Replace the battery (3volt)
Repair the Digi B/D
Yes
Yes
Periodic Noises
All alarm has been
turned off
Display lead fault
Message appears
No
The printer function abnormally
No
Turn functions not in use off
(Mode display set para)
Replace the printer