atient’s Cable
⚫
Clean the patient cables with a gauze pad moistened with a soap solution.
⚫
To disinfect patient cables, wipe the cables with a gauze moistened with diluted alcohol
or a glutaraldehyde-based dis-infectant.
⚫
Ethylene oxide is suitable for intensive disinfection (almost sterilization), but it shows that
the service life of cables and lead wires is reduced.
⚫
Dry thoroughly with a lint-free cloth.
CAUTION
Do not use disinfectants that contain phenol as they can spot plastics. Do not autoclave or
clean accessories with strong aromatic, chlorinated, ketone, ether, or ester solvents. Never
immerse electrical connectors.
When cleaning, do not apply excessive pressure or bend the cable unnecessarily. Excessive
pressure can damage the cable.
Reusable ECG Electrodes
Clean the electrode cup regularly with a toothbrush. When removing gel-like residues, use a soft
brush with flowing water. Wipe the electrode with a soapy cloth moistened with soapy water.
⚫
Sterilize the electrode by soaking the diluted alcohol in cloth.
⚫
Dry thoroughly with a lint-free cloth.
Reusable SpO2 sensor
Clean the SpO2 sensor by wiping it with soapy water gauze. Disinfect the sensor by wiping with
70% alcohol solution. Allow the sensor to dry completely with a lint-free cloth before applying to
the animal.
Capnostat sensor
Wipe the sensor surface with a damp cloth. Do not attempt to wet the sensor or disinfect it with
hot water. Allow to dry completely with a lint-free cloth
Reusable Temperature probes and cables
Do not use excessive pressure or flex the cables as this can stretch the covering and break the
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...