94
BM5Vet Pro
User’s Manual
SMALL: 60-140mmHg
MEDIUM: 80- 140mmHg
LARGE: 120-250mmHg
A-4.
SETTING TIME
How to apply pressure value setting.
Once: When the blood pressure is
measured for the first time, the
pressure is set to the set pressure
value, but automatically adjusted
according to the animal's blood
pressure value.
Every Time: Whenever blood pressure
is measured, pressurize to the set
pressure value every time
Once,
Every Time
A-5.
Measure Interval
A menu to set Interval time when the
blood pressure is measured.
After setting INTERVAL, you must press
NIBP KEY to start NIBP measurement.
1min, 2, 3, 4, 5, 10,
15, 20, 30, 60, 90, 2
hours, 4, 8.
A-6.
VITAL SIGN REVIEW
Record the last 40 blood pressure
readings
Warning
Check periodically to see if the circulation from the cuff to the distal part of the animal's
arm is good.
1minute and 2minute intervals when using automatic measurement, check the animal's
condition frequently. It is not recommended for measuring blood pressure for a long time if
the measurement time period is set to 10minutes or less.
When changing the cuff, be sure to adjust the actual size of the cuff and the size of the cuff
set on the equipment.
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...