144
BM5Vet Pro
User’s Manual
Caution
The calibration gas must be used in the concentration (%) range shown below
◼
Nitrous oxide (N2O)> 30%
◼
Carbon dioxide (CO2)> 2.5%
◼
Oxygen: O2> 40%
◼
Agents> 1%
Caution
The calibration gas must use gas with balance N2. (Eg 21% O 2, bal. N 2). After the
calibration is finished, check the measurement accuracy of the BMGA using a calibration gas
of another component