IEC 61000-4-11
<5% U
т
(>95% dip in U
т)
for 0.5cycle
40% U
т
(60% dip in U
т)
for 5 cycle
70% U
т
(30% dip in U
т)
for 25 cycle
<5% U
т
(<95% dip in U
т)
for 5 s
<5% U
т
(>95% dip in U
т
)
for 0.5cycle
40% U
т
(60% dip in U
т)
for 5 cycle
70% U
т
(30% dip in U
т)
for 25 cycle
<5% U
т
(<95% dip in U
т
)
for 5 s
Mains power quality should
be that of a typical comme
rcial or hospital environmen
t. If the user of the BM5
Vet Pro system requires cont
inued operation during pow
er mains interruptions, it is
recommended that the BM5
Vet Pro system be powered
from an uninterruptible pow
er supply
or a battery
Note:
U
т is the a.c. mains voltage prior to application of the test level.