206
BM5Vet Pro
User’s Manual
AG2E-DES
●
●
AG2I-DES
●
●
AG2E-ENF
●
●
AG2I-ENF
●
●
AG2E-HAL
●
●
AG2I-HAL
●
●
AG2E-ISO
●
●
AG2I-ISO
●
●
AG2E-SEV
●
●
AG2I-SEV
●
●
Biosignal Alarm Limits and Min/Max setting Value
Biosignal
Class
Animal
Type
Biosignal
Alarm
Limits
Min./Max.
Position
(IBP
ONLY)
Alarm
Lower
Upper
Min.
Max.
ECG
PUPPY
HR
90.0
180.0
15.0
350.0
ST
-0.4
0.4
-2.0
2.0
PVC
0.0
20.0
0.0
99.0
CAT
HR
90.0
200.0
15.0
300.0
ST
-0.4
0.4
-2.0
2.0
PVC
0.0
20.0
0.0
99.0
DOG
HR
60.0
150.0
15.0
300.0
ST
-0.4
0.4
-2.0
2.0
PVC
0.0
20.0
0.0
99.0
HORS
E
HR
60.0
150.0
15.0
300.0
ST
-0.4
0.4
-2.0
2.0