BM5Vet Pro
User’s Manual
carefully look at the SpO2 waveform and check the probe position in the animal.
SPO
2
Waveform with Artifact
Stability of SPO
2
Values
The stability of the displayed SPO
2
values can also be used as an indication of signal validity.
Messages are provided in the SPO
2
values window to aid you in successful SPO
2
monitoring.
WARNING
In the monitoring of animal, it is possible that adverse conditions may lead to a disturbed
signal going unnoticed. In this situation, artifacts can simulate a plausible parameter
reading, so that the monitor fails to sound an alarm. In order to ensure reliable animal
monitoring, the proper application of the probe and the signal quality must be checked at
regular intervals.