175
BM5Vet Pro
User’s Manual
internal wires.
⚫
Clean the probes with a 3% hydrogen peroxide or 70% alcohol.
⚫
Quickly immerse the cables in a detergent solution.
⚫
Make sure the probe’s tip is firmly connected.
CAUTION
Never boil or autoclave the cable. Vinyl withstands temperatures up to 100
°
C but begins to
soften at around 90
°
C. Handle gently when hot and wipe away from the tip toward the
cable.
CAUTION
Decisions on disinfection should be made by the user organization in accordance with the
integrity of the wires or lead wires.
Note
The equipment should be inspected regularly once a year. For inspection items, refer to the
user manual or service manual.
Carefully inspect the main unit and sensor after cleaning the equipment. Do not use damaged or
old equipment.
Clean the exterior of the equipment at least once a month using a soft cloth moistened with
lukewarm water or alcohol. Do not use laquers, thinners, ethylene, or oxidizers that could damage
the equipment.
Make sure that the cables and accessories are free from dust and dirt, then wipe them with a soft
cloth moistened with 40 ° C water. Please wipe it with clinical alcohol at least once a week.
Do not immerse the accessory in liquid or detergent. Also, make sure that no liquid penetrates
the instrument or probe.
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...