190
BM5Vet Pro
User’s Manual
IBP Performance (Option)
Channels
2
Measurement range
-50 to 300mmHg
Accuracy
<100mmHg :
±
1mmHg
>=100mmHg :
±
1% of reading
Pulse rate measurement
range
0 to 300bpm
Zero balancing
Range :
±
200mmHg
Accuracy :
±
1mmHg
Drift :
±
1mmHg over 24hours
Transducer sensitivity
5µV/mmHg
Pulse rate measurement
range
0 to 300bpm
Sidestream CO2 (Option)
Measurement range
0 to 150 mmHg, 0 to 19%
Accuracy
0-40mmHg
±
2 mmHg,
41-70mmHg
±
5% of reading
71-100mmHg
±
8% of reading,
101-150mmHg
±
10% of reading
Respiration rate
2 to 150 breath per minute
Respiration accuracy
±
1breath per minute
Mainstream CO2 (Option)
Measurement range
0 to 150 mmHg, 0 to 19%
Accuracy
0-40mmHg
±
2 mmHg,
41-70mmHg
±
5% of reading
71-100mmHg
±
8% of reading,
101-150mmHg
±
10% of reading
Respiration rate
0 to 150 breath per minute
Respiration accuracy
±
1breath per minute
Anesthetic GAS/O2 - Phasein (Option)
Method
Infra-red absorption characteristic
Gas
CO2, O2, N2O, Des, Iso, Enf, Hal, Sev
Warm-up time
Mainstream (IRMA AX+)
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...